ETRI-Knowledge Sharing Plaform

KOREAN
논문 검색
Type SCI
Year ~ Keyword

Detail

Journal Article Sensing probe for 3–6 GHz microwave imaging systems
Cited 9 time in scopus Share share facebook twitter linkedin kakaostory
Authors
Kwang-Jae Lee, Jang-Yeol Kim, Seong-Ho Son, Jongmoon Lee, Soonik Jeon
Issue Date
2014-07
Citation
Electronics Letters, v.50, no.15, pp.1049-1050
ISSN
0013-5194
Publisher
IET
Language
English
Type
Journal Article
DOI
https://dx.doi.org/10.1049/el.2014.1923
Abstract
A sensing probe for medical imaging diagnostic systems is presented. The proposed probe consists of a double-ridged waveguide (DRWG) filled with dielectric ceramic and an aperture with a tapered ridge. The DRWG is used to suppress higher-order mode radiations in the rectangular waveguide. The proposed probe is designed for broadband sensing in an immersion liquid with a complex dielectric constant. The performance of the fabricated probe successfully covers the frequency band of 3-6 GHz. The probe suppresses higher-order modes and provides symmetric and similar radiation patterns within the operating frequency range. © 2014 The Institution of Engineering and Technology.
KSP Keywords
6 GHz, Complex Dielectric Constant, Dielectric ceramic, Medical Imaging, Microwave Imaging, Rectangular Waveguide, Sensing probe, broadband sensing, diagnostic system, frequency band, higher-order mode