학술대회 Experimental Measurements of Multi-Carrier based Doppler Heartrate Tracker
Cited 2 time in scopus Download 4 time Share share facebook twitter linkedin kakaostory
저자
김동규, 최영우, 허재두
발행일
201610
출처
Biomedical Circuits and Systems Conference (BioCAS) 2016, pp.184-187
DOI
https://dx.doi.org/10.1109/BioCAS.2016.7833762
협약과제
16MH3300, 폐암, 치매의 생체신호 및 매몰자 탐지를 위한 융합원천 기술개발, 허재두
초록
This study presents experimental results of multi-carrier based Doppler radar signal processing for vital signs tracking. When small motion-transients exist, it is expected that a disturbance at a carrier frequency is different from other carrier frequencies. The multi-carrier based Doppler radar shows the possibility to overcome the motion-transient effect in some situations. Six tests are performed to measure multi-carrier Doppler signals. Their heartrate tracking results show that the mean values of the tracked heartrates are within plus/minus 3% of the ground truth R-R interval frequency of their electrocardiogram measurements. It is also shown that the variance values of the tracked heartrates are all within 0.0062 Hz. The proposed signal processing scheme seems to have potential to track vital signs while small motion-transients exist.