ETRI-Knowledge Sharing Plaform

KOREAN
논문 검색
Type SCI
Year ~ Keyword

Detail

Conference Paper Comparison of Conductive Fabric Sensor and Ag-AgCl Sensor Under Motion Artifacts
Cited 13 time in scopus Share share facebook twitter linkedin kakaostory
Authors
I.B.Lee, S.C.Shin, Y.W.Jang, Y.S.Song, J.-W.Jeong, S.Kim
Issue Date
2008-08
Citation
International Conference of the IEEE Engineering in Medicine and Biology Society (EMBS) 2008, pp.1300-1303
Publisher
IEEE
Language
English
Type
Conference Paper
DOI
https://dx.doi.org/10.1109/iembs.2008.4649402
Abstract
A wearable electrocardiogram(ECG) device using conductive fabric sensor was compared with traditional Ag-AgCl electrode ECG device. The ECG signals were measured under existence of motion artifacts on variable running speed using treadmill to verify that wearable device can substitute traditional ECG device. A signal to noise ratio (SNR) and RR interval were compared between the two devices. The SNR of wearable device was similar or higher than that of clinical device and difference of RR interval was 2ms. The results show that the wearable ECG device using conductive fabric sensor can make similar performance with ECG device using Ag-AgCl electrode even under motion artifacts.
KSP Keywords
Ag-AgCl, ECG device, ECG signal, Fabric sensor, Motion artifact, RR interval, Running speed, Signal noise ratio(SNR), Signal-to-Noise, Wearable ECG, Wearable device