Journal Article Performance Study of the Wearable One-Lead Wireless Electrocardiographic Monitoring System
Cited 11 time in scopus Share share facebook twitter linkedin kakaostory
Authors
Sung Youp Hong, Young Mo Yang, Seung Hwan Kim, Seung Cheol Shin, In Bum Lee, Yong Won Jang, Ki Seong Kim, Nak Jin Seong, Hwa Yeon Yi, Jang Young Lee
Issue Date
2009-03
Citation
Telemedicine and e-Health, v.15, no.2, pp.166-175
ISSN
1530-5627
Publisher
Mary Ann Liebert
Language
English
Type
Journal Article
DOI
https://dx.doi.org/10.1089/tmj.2008.0071
Abstract
This study attempts to compare and assess the performance of a wearable electrocardiogram (ECG) using a sensing fabric electrode and a Bluetooth짰 network with a conventional ECG. A one-lead ECG examination was performed using Bioshirt and an iWorx짰 214 while walking or running at 3, 6, and 9 km per hour. A correlation coefficient of a heart rate variability (HRV) between these two devices was higher than 0.96 and power spectral density of HRV measured also showed an excellent agreement. Thus, both of these two ECG devices showed similar detection capability for R peaks. The measured values for wave duration and intervals of both devices concur with each other. The intensity of noise is controversial. The ECG device using a sensing fabric electrode and a wireless network showed an ECG signal detection and transmission capability similar to that of a conventional ECG device. © Copyright 2009, Mary Ann Liebert, Inc.