Journal Article Stabilization Effect by the Temperature of Immersion Liquid in the 3-6 GHz Microwave Tomography System
Cited 0 time in scopus Share share facebook twitter linkedin kakaostory
Authors
Jang-Yeol Kim, Kwang-Jae Lee, Seong-Ho Son, Hyuk-Je Kim, Jong-Moon Lee, Soon-Ik Jeon, Nam Kim
Issue Date
2016-02
Citation
Microwave and Optical Technology Letters, v.58, no.2, pp.461-464
ISSN
0895-2477
Publisher
John Wiley & Sons
Language
English
Type
Journal Article
DOI
https://dx.doi.org/10.1002/mop.29587
Abstract
A study on temperature stabilization effect of the immersion liquid in a 3-6 GHz microwave tomography (MT) system for medical diagnosis is presented. The designed temperature control device unit is applied in the MT system for the analysis of the temperature impact. This control unit with an advanced control method is required to precisely maintain the temperature of the immersion liquid for obtaining stable measurement data. If the temperature of the liquid is maintained ideally, the peak to peak deviation of the measured amplitude and phase are very small within approximately 0.18 dB and 1.3 degree over the entire frequency range. The temperature variation of the liquid is a key factor for high spatial resolution to detect small tumors. The detailed results are presented and discussed.
KSP Keywords
6 GHz, Advanced control, Control Unit, Control device, Frequency range, Key factors, MT system, Medical diagnosis, Microwave tomography, Stabilization effect, Temperature Control