Conference Paper Noninvasive glucose monitoring evaluation with diffused transmitted NIRS in soft silicone-based tissue phantoms
Cited 1 time in scopus Share share facebook twitter linkedin kakaostory
Authors
Jongdeog Kim, Bong Kyu Kim, Chul Huh
Issue Date
2023-10
Citation
IEEE Sensors 2023, pp.1-4
Publisher
IEEE
Language
English
Type
Conference Paper
DOI
https://dx.doi.org/10.1109/SENSORS56945.2023.10325071
Abstract
Solid tissue phantoms have been developed based on soft silicone elastomers with hydrophile additives, such as absorptive dyestuffs and glucose concentrations. For two prototypes of phantoms, we investigated the relative transmittance and extinction characteristics with transmitted near-infrared spectroscopy (NIRS) using an optical probe sensor. Notwithstanding the surface reflection and multiple scattering in solid tissue phantoms, we have successfully verified the performance of noninvasive glucose detection using a proposed normalized extinction difference method. The linear regression analysis for prediction glucose levels has shown a 0.99 adjusted coefficient of determination to reference glucose concentrations from 0 to 987 mg/dL.
KSP Keywords